The MegaPulse DF-80PF DSW Surge Tester is designed to conduct the test
shown in AAMI DF80, App. DD, Table US.DD.1 and Figure US.DD.1 - Damped
sinusoidal waveform parameters. The resulting waveform is in
accordance with those requirements. For waveform and
tolerance/compliance graphs and information, click here.
The DF-80PF DSW is built with 50 ohms output impedance. If another impedance is desired, please contact us.
The DF-80PF DSW is designed as a laboratory instrument with a duty cycle 1 pulse every 5 minutes continuous within tolerances. Other
duty cycles are optionally available:
Option 60. Allows one test every 60 sec continuously.
Option 12. Allows one test every 12 sec continuously.
The power output of the DF-80PF DSW is 360J. A front panel switch
reverses the output polarity.
Positive and negative pulses can be delivered manually using the front
panel buttons, or the pulses can be set to operate under computer
control using the TestMinder Megapulse PF option.
Actual output waveforms are available for review in the Manuals section below.
MegaPulse DF-80PF DSW ships with cables, graphs of theoretical and actual
waveforms, and a Calibration Certificate. One-year warranty.