MegaPulse
Biphasic PF 360/100J:

SKU: 00-MPBiphasic PF 360/100J

Surge Tester Biphasic PF 360/100J

Call for Price

The Biphasic PF 360J and 100J outputs the pulse used in defibrillators with a built in 50Ω patient load, the MegaPulse can recharge and trigger the next pulse within 30 seconds.

Designed & Built in the USA

  • 1-Year Warranty
  • 100% Secure Checkout
Guaranteed Safe Checkout

The MegaPulse 360J/100J can be used to test products in research and production environments. The pulse energy can be switched between 360J and 100J.  See waveforms in the Manuals Section below.  The biphasic output pulse mimics the newest defibrillator output standard but allows more repeatability and faster pulsing than a defibrillator would.

The output is stable and provides the required 360J power (and 100J when required) in a production test or continuous test environment. The available autotrigger control, TestMinder P, provides very repeatable outputs which makes testing results more meaningful.

The MegaPulse 360J/100J is used to prove any part which would be subjected to defibrillator pulses.

The biphasic output reversal timing can be configured at the factory for any timing desired within the pulse. Output power is also configurable upon request.

Output voltage is approximately 2kV, with slight variances to meet the 360J power requirement. Output may be as high as 2040V in order to provide the power required. Output voltage is less when 100J output is selected, see waveforms in Manuals section below.

With the available 00-MPH Monophasic waveform option, the MegaPulse 360J can also output the truncated isolation waveform of AAMI DF-1, Figure A.2 for additional testing of defibrillator leads.

May be configured with TestMinder P (optional).

WIth the configurable T1 and T2, and the monophasic option, this tester is very customizable.  Please call or write with any questions or thoughts.

Features

• 60J or 100J selector on front panel.
• Digital output voltage knob on the front panel.
• Integrated High Voltage on hold circuit.
• External interlock disables HV output and defeats test when circuit is open.
• Available USB TestMinder, computer control with auto-trigger via USB.
• Front panel indicates internal HV and polarity.
• Polarity button selector on the front panel.
• Microcontroller technologies.
• Leads, manual and calibration certificate included.
• One year calibration cycle
• One year warranty.

Specifications

TESTING CIRCUITRY
Output Peak Voltage: 2000V ±3%.
Main Capacitance: 200uF ±10%.
Energy: 100J / 360J ±10%.
Tphase1: (6.8ms – 7.9ms).
Tphase2: (4.5ms – 5.3ms).
Commutation Time: <2ms.
Tilt: (63.9 – 71).
Internal Patient Resistance: 50Ω ±5%.
Output Impedance: 25Ω ±5%.
Voltage Control: Manually, by turning a knob in the front panel or via USB.
Polarity Control: Positive and Negative. Alternating control available with USB.
Voltage Display: 4 Digit LED Display.
Voltage Meter resolution: 5V.
Voltage Meter Accuracy: ±3V.
Duty Cycle: 30s between pulses.

ELECTRICAL
Line Voltage: 120V AC, 50/60Hz.
230V AC ±10%, 50 Hz ±1 Hz (CE for EU).

SAFETY
Safety Interlock 2-pin terminal block on rear panel; shorted for operation.

OUTPUTS – FRONT
OUTPUT: High Voltage insulated rigid 4mm socket – Red.
RETURN: High Voltage insulated rigid 4mm socket – Black.

OUTPUTS – REAR
TESTMNINDER: DB9 Female (optional).
Inlet Module: Inlet module with fuse and power switch combination.
Relay Interface: Terminal block with screw.