• Defib-proof and Energy Reduction Tester 10KV
  • Product Views
Datasheet

MegaPulse D5-PF-001:
Defib-proof and Energy Reduction Tester 10KV

Call for Price

The most advanced medical surge tester available, the D5-PF is built to provide defib-proof testing to the requirements of:

  • IEC 60601-1:2005 (Issue 3) Figure 9 and 10 Defib-proof Differential- and Common-Mode Test  (IEC 60601-1 Figure 50 in older versions).
  • IEC 60601-1:2006 Figure 11 Energy Reduction Test;
  • IEC 60601-2-4: 2010 Figure 201.110 Overload Test (optional)
  • IEC 60601-2-34 Issue 2 (optional);
  • IEC 60601-2-34 Issue 3 (included);
  • IEC 60601-2-49;
  • EC-13;
  • EC-53;
  • IEC 60601-2-25:2011;
  • IEC 60601-2-27:2011.

This is an IEC 60601 Defibrillation-proof tester.  For the Energy Reduction test, joules consumed in the 100Ω resistor are immediately displayed after each pulse. The MegaPulse D5-PF tests protection against the effects of defibrillation (Common Mode, Differential Mode) and Energy reduction test (EC-13) and as noted in the Standards shown above. The MegaPulse D5-PF outputs a 360J Defib-proof test pulse per IEC 60601-1 for the Energy Reduction Test.  It can also be used as a Defibrillator Lead Tester.  

Detailed Information about the MegaPulse D5-PF

White Paper - Oscilloscope Setup to Capture Y1-Y2 waveform in accordance with Iss. 3 Fig. 10

Listen to Expert Highlights


Available Options

Reviews (0)

Write a Review

Note: HTML is not translated!
    Bad           Good

Features

  • Direct Joules readout on front panel after Energy Reduction Test.
  • Increased accuracy in accordance with IEC 60601-2-25 and -2-27:2011 requirements.
  • 20 second period between pulse delivery,
  • Increased signal generator voltage for -2-27 tests (Order Option SG).
  • Integrated Pass/Fail reference for checking the equipment and test setup.
  • Dry Capacitor delivers >2.5M pulses at 20 second intervals.
  • Integrated energy measurement, see demonstration video.
  • USB Testminder, Computer Control.
  • External interlock, disables HV output and defeats test when circuit is open.
  • Ergonomically designed for safety, speed and efficiency.
  • Front panel indicates polarity
  • Output voltage and polarity are controlled manually on front panel or via TestLink.
  • Cables, manual and calibration certificate included.
  • One year calibration cycle.
  • One year warranty.

Specifications

Output Voltage0-10KV ±1%
Output Voltage0-10KV ±1%
Main Capacitance32 uF ±5%
Inductances500uH and 25 mH ±10%
Main Resistances100 Ω, 50 Ω and 400 Ω ±1% non inductive
Voltage ControlManually, by turning a knob in the front panel
Polarity controlPositive and Negative. Alternating control available with optional computer control
Polarity ControlPositive an Negative, by front panel push button
Duty Cycle1 Pulse every 20 Seconds *optional 12 Seconds
Voltage Display4 Digits LED Display
Voltage Resolution5 Volt
Line Voltage120 VAC AC, 50/60 Hz *optional different line voltages available (100V, 110V, 220V, 230V, 240V)
Operating Temperature15-40 ºC
Relative Humidity Range0-90% non-condensing
Dimensions17" wide x 18" high x 17" in deep

Manuals

Download Manual
MegaPulse D5-PF rev 3.pdf Download Manual
Waveforms/IEC 60601-1 Figure 10 simulation.pdf Download Manual
Waveforms/IEC 60601-1 Fig. 10 Voltage Duration Waveform calculated.pdf Download Manual
Waveforms/IEC 60601-1 Fig. 10 Voltage Peak Waveform actual.pdf Download Manual
Waveforms/IEC 60601-1 Fig. 10 Voltage Rise Waveform calculated.pdf Download Manual
Waveforms/IEC 60601-1 Fig. 10 Voltage Rise Waveform actual.pdf Download Manual
Waveforms/IEC 60601-1 Fig. 10 Current Duration Waveform calculated.pdf Download Manual
Waveforms/IEC 60601-1 Fig. 10 Current Rise Time Waveform calculated.pdf Download Manual
Waveforms/EC13 Voltage Duration Waveform calculated.pdf Download Manual
Waveforms/EC13 Voltage Duration Waveform actual.pdf Download Manual
Waveforms/EC13 Voltage Rise Waveform actual.pdf Download Manual
Waveforms/EC13 Current Duration Waveform calculated.pdf Download Manual
Waveforms/EC13 Voltage Duration 100 Ohm Load Waveform calculated.pdf Download Manual
Waveforms/EC13 Voltage Duration 50 Ohm Load Waveform calculated.pdf Download Manual
Waveforms/EC13 Voltage Duration 25 Ohm Load Waveform calculated.pdf Download Manual

Contact Us

Corporate Sales Office (USA)

Compliance West USA
650 Gateway Center Way Suite D
San Diego, CA 92102 USA

US Representatives

Product Update Bulletin

Newest Product: HT-3500S RF, "ruggedized" hipot tester

The HT-3500S RF tester is specifically designed for the challenges common to operating in the field. Along with heavy duty internal brackets and a sturdy, shock-absorbing case, this unit includes adjustable voltage output, voltage ramp, test time, leakage current and ground continuity limit settings - all controlled from an easy-to-access front panel. Call for pricing.

Read More

Available: 16A option for CDT-240

Perform one of the most difficult tests required with both speed and accuracy. The CDT-240 now offers a 16A option along with designed presets to implement the plug discharge test as described in IEC 61010, 60065 and 60335.

Read More

MegaPulse Defib 5P reaches EOL - Replaced by D5-PF

Compliance West USA is announcing the EOL (End-of-LIfe) of our MegaPulse Defib 5P series effective 07/14/2015. This unit has been replaced by our MegaPulse D5-PF advanced tester. Compliance West will continue to support, repair and calibrate the Defib 5P throughout the terms of any existing warranty agreements or up until the product's End-of-Support date. Please contact us with any questions.

Read More