Function Tester, Ground (Pass 0.025, Fail 0.1 ohm) Leakage (Pass 848k Fail 424kohm) Arcover 750V ± 400V

  • Function Tester, Ground (Pass 0.025, Fail 0.1 ohm) Leakage (Pass 848k Fail 424kohm)  Arcover 750V ± 400V
  • Product Views
  • Function Tester, Ground (Pass 0.025, Fail 0.1 ohm) Leakage (Pass 848k Fail 424kohm)  Arcover 750V ± 400V
Datasheet

HTT-1R MedmA:
Function Tester, Ground (Pass 0.025, Fail 0.1 ohm) Leakage (Pass 848k Fail 424kohm) Arcover 750V ± 400V

  • $425.00

The HTT-1R provides a fast, safe and reliable way to verify proper operation of Dielectric Withstand (hipot) and Ground Continuity/Ground Bond Testers between calibration cycles, in accordance with requirements of TUV and other European agencies.  It uses actual loads to simulate barely passing and barely failing results to exercise the tester(s), and is designed to be used as a verification test fixture in compliance with the requirements for shift by shift verification stated in ECS CIG-023 Inspection Report, Section 4. See the Manuals Section below for an example of the CIG-023 Inspection Report.

It allows the operator to check the result of the three failure modes encountered in production hipot and ground circuit testing (open ground, high leakage current, and dielectric breakdown) separately or in any combination to verify proper test equipment operation. A passing test is also simulated.

The test leads allows connection of the HTT-1R to any hipot tester using a standard NEMA 15A receptacle, or use the leads on the hipot and ground tester to connect directly to the HTT-1R. The manual contains anticipated results and suggests testing procedures. The maximum voltage the HTT-1R can test is 3000Vac/4200Vdc.

The HTT-1R Family has ranges suitable for many scenarios, and a Custom solution. 

HTT-1R MedmA, Function Checker, Ground (Pass 0.025, Fail 0.1Ω) Leakage (Pass 848k Fail 424kohm) Arcover 750V ± 400V.

Reviews (0)

Write a Review

Note: HTML is not translated!
    Bad           Good

Features

  • The HTT-1R Medma simulates the following failures:
  • Ground Continuity Chassis to Ground (0.01Ω), and/or
  • High Leakage Line to Ground (424kΩ), and/or
  • Dielectric Withstand Failure (Arc View window displays an arc at 750Vac ± 400V).
  • The HTT-1R can also be set to simulate passing tests:
  • Ground Continuity Chassis to Ground (0.025Ω), and/or
  • Normal Leakage Line to Ground (848kΩ), and /or
  • Dielectric Withstand OK (up to 3000 volts)
  • Arc Testing is at 750 ± 400Vac. Results visible through Front Panel Window.
  • Works with any manufacturer's Dielectric Withstand or Ground Continuity Tester. The HTT-1R simulates the failures and your tester should return results consistent with the switch settings on the front panel of the HTT-1R
  • The HTT-1R is a passive device. No batteries or power is required.
  • Verifies proper operation of your equipment in less than a minute.
  • Portable.
  • Light weight.
  • Simple to operate.
  • Saves time.
  • For testing of Ground Bond Testers up to 30A..
  • 1 year warranty.

Specifications

For Specifications, see Datasheet Link, above.
For more Specifications, see Datasheet Link, above.
Max Current capability of Ground Test circuit30A
Setting of PASS for Leakage Current 848kΩ
Setting of FAIL for Leakage Current424kΩ
Max Voltage capability of Hipot Test Circuit3000Vac / 4200Vdc
Setting of PASS for Ground Circuit 0.025Ω
Setting of FAIL for Ground Circuit 0.1Ω
Setting of Arc Detection CircuitArc-over at 750V +/- 400Vac / 1050V +/- 560Vdc
ENVIRONMENTAL
Operation Temp15 - 40 C
Relative Humidity Range0-90% non-condensing
GENERAL
Dimms5.3"(W) x 6.8"(H) x 2.3"(D)
Wt2lbs

Manuals

Download Manual
HTT-1R.pdf Download Manual
PD CIG 023 - September 2014.pdf Download Manual

Contact Us

Corporate Sales Office (USA)

Compliance West USA
650 Gateway Center Way Suite D
San Diego, CA 92102 USA

US Representatives

Product Update Bulletin

Newest Product: HT-3500S RF, "ruggedized" hipot tester

The HT-3500S RF tester is specifically designed for the challenges common to operating in the field. Along with heavy duty internal brackets and a sturdy, shock-absorbing case, this unit includes adjustable voltage output, voltage ramp, test time, leakage current and ground continuity limit settings - all controlled from an easy-to-access front panel. Call for pricing.

Read More

Available: 16A option for CDT-240

Perform one of the most difficult tests required with both speed and accuracy. The CDT-240 now offers a 16A option along with designed presets to implement the plug discharge test as described in IEC 61010, 60065 and 60335.

Read More

MegaPulse Defib 5P reaches EOL - Replaced by D5-PF

Compliance West USA is announcing the EOL (End-of-LIfe) of our MegaPulse Defib 5P series effective 07/14/2015. This unit has been replaced by our MegaPulse D5-PF advanced tester. Compliance West will continue to support, repair and calibrate the Defib 5P throughout the terms of any existing warranty agreements or up until the product's End-of-Support date. Please contact us with any questions.

Read More