Compliance West USA Hipot Testers Header

MegaPulse DF-1P Current Carrying

DF-1 Image

The DF-1P megapulse Tester is designed to generate the waveform described in AAMI DF-1, Figure B.2., for the connector cavity current-carrying test in AAMI DF-1, Annex B. The test can be set to automatically conduct the first 490 cycles (500 ± 50 as show in standar)of the test, and then stop testing to allow laboratory personnel monitor the last 10 cycles in accordance with the Standard. Voltage and current BNC outputs make waveform analysis convenient. For example, please take a look at the output voltage and current waveforms.

Our TestMinder computer control is included to allow automated repetitive testing.

A front-panel switch reverses the output polarity. Additionally provided with BNC oscilloscope output for voltage (100:1) and current (1V/A) for output monitoring.

The waveform output is 1.05uSec rise time, with a 18mS duration, with an output impedance of 15 ohms.

MegaPulse DF-1P Current carrying Tester ships complete, ready to perform the DF-1 test:

  • TestMinder to conduct automated repetitive testing
  • BNC Current Jack (1V/mA)
  • BNC Voltage Jack (100:1)
  • Cables
  • Graphs of theoretical and actual waveforms
  • Calibration Certificate.
  • One-year warranty.

MegaPulse DF-1P Current carrying Tester.

    TOP

Maximum Voltage output 1300 V
Current Duration   18 mS
Meter accuracy   ±5 %
Operation temperature 15-40 °C
Relativity humidity range 0-90 % Non condensing
Standards AAMI DF-1 App B
Figure B.2
 
 
 
 
 
 
Storage   No
Product weight   35 Lbs
Warranty   1 Year
Calibration   1 year
ECCN number   3B992
HTS/Schedule B   9030.39.0100
Product Dimensions   17 w x 17 L x 6 H
PLC interface   Optional
Autotriger   Optional
Continuos Full voltage Optional
Testminder   Optional
Energy calculation   Optional
Power supply voltage 120 VAC 50/60 Hz
what is in the box Megapulse DF-1P Current carrying
2 high voltage Test leads
18 AWG Line power cord
Instruction manual
 
Interfaces I/O  USB
Price   Call

 

    TOP

Option: 240-240V Mains power (requires 3U enclosure)
Option: IG-Optional IGBT Switch
Option MPH: Monophasic 200uF waveform addition...
Option P6015: High accuracy Tektronix Probe for proper waveform measurement

    TOP
Generates the waveform required for the Current carrying Test, AAMI DF-1, Appendix B, and described in AAMI DF-1, Figure B.2.
Designed with operator safety in mind: Set up the EUT to the right and behind the tester, and a right-handed operator will find himself standing to the left of the tester, allowing maximum separation.
Theoretical graphs and actual graphs showing the output of your tester are included with each MegaPulse DF-1P, allowing easy sanity checks and proof to visiting agency engineers of test setups.
BNC Oscilloscope outputs allows quick connection and easy evaluation of voltage and current output waveforms.
Reverse polarity waveforms available by pushing a button on the front panel. No need to change the test setup. Positive polarity indication minimizes errors.
TestMinder computer control allows automation of the required 490 cycles; then the operator can conduct the last 10 cycles in which the current waveform must be monitored. (Waveform capture is not provided).
1-year calibration cycle.
All test leads included.
1 year warranty

    TOP