Compliance West USA Hipot Testers Header

MegaPulse DF-1 Lead Connector Tester

DF-1 Image

The DF-1 Cutoff Tester is designed to generate the waveform described in AAMI DF-1, Figure A.2., for the Lead Connector electrical isolation test in AAMI DF-1, Annex A. The test can be set to automatically conduct the first 490 cycles of the test, and then stop testing to allow laboratory personnel monitor the last 10 cycles in accordance with the Standard. Voltage and current BNC outputs make waveform analysis convenient. For example, please take a look at the output voltage and current waveforms of a simulated test. (Please note that these tests were conducted at 1300 Volts instead of the required 1500 volts. The MegaPulse DF-1 is capable of a pulse output of up to 2500 volts.)

Our TestMinder computer control is included to allow automated repetitive testing.

The test signal includes the waveform cutoff at 750V/18msec as shown in Figure A.2. Please see the Standard Figure A.2, and the actual output of the MegaPulse DF-1 Lead Connector Tester.

A front-panel switch reverses the output polarity. Additionally provided with BNC oscilloscope output for voltage (100:1) and current (1V/mA) for output monitoring.

The waveform output is 1.5uSec rise time, with a 18mS duration, with an output impedance of 500 ohms.

Click for a view of the outputs showing the Current and Voltage BNC jacks.

MegaPulse DF-1 Lead Connector Tester ships complete, ready to perform the DF-1 test:

  • TestMinder to conduct automated repetitive testing
  • Current cutoff and waveshape truncation
  • BNC Current Jack (1V/mA)
  • BNC Voltage Jack (100:1)
  • Cables
  • Graphs of theoretical and actual waveforms
  • Calibration Certificate.
  • One-year warranty.

MegaPulse DF-1 Lead Connector Tester.

    TOP

Use

Designed to generate the waveform described in AAMI DF-1, Figure A.2., for the Lead Connector electrical isolation test in AAMI DF-1, Annex A.

    TOP

Option: 240-240V Mains power (requires 3U enclosure)
Option: IG-Optional IGBT Switch
Option IFD:Insulation Fault Detection.
Option MPH: Monophasic 200uF waveform addition...
Option P6015: High accuracy Tektronix Probe for proper waveform measurement

    TOP
Generates the waveform required for the Lead Connector Test, AAMI DF-1, Appendix A, and described in AAMI DF-1, Figure A.2. The waveform is truncated at 50% of peak value and 18mSec as required.
Designed with operator safety in mind: Set up the EUT to the right and behind the tester, and a right-handed operator will find himself standing to the left of the tester, allowing maximum separation.
Theoretical graphs and actual graphs showing the output of your tester are included with each MegaPulse DF-1, allowing easy sanity checks and proof to visiting agency engineers of test setups.
BNC Oscilloscope outputs allows quick connection and easy evaluation of voltage and current output waveforms.
Reverse polarity waveforms available by pushing a button on the front panel. No need to change the test setup. Positive polarity indication minimizes errors.
TestMinder computer control allows automation of the required 490 cycles; then the operator can conduct the last 10 cycles in which the current waveform must be monitored. (Waveform capture is not provided).
1-year calibration cycle.
All test leads included.
1 year warranty

    TOP

Ticket info - call 800-555-1212