The DF-1 Cutoff Tester is designed to generate the waveform described in AAMI DF-1, Figure A.2., for the Lead Connector electrical isolation test in AAMI DF-1, Annex A. The test can be set to automatically conduct the first 490 cycles of the test, and then stop testing to allow laboratory personnel monitor the last 10 cycles in accordance with the Standard. Voltage and current BNC outputs make waveform analysis convenient. For example, please take a look at the output voltage and current waveforms of a simulated test. (Please note that these tests were conducted at 1300 Volts instead of the required 1500 volts. The MegaPulse DF-1 is capable of a pulse output of up to 2500 volts.)
Our TestMinder computer control is included to allow automated repetitive testing.
The test signal includes the waveform cutoff at 750V/18msec as shown in Figure A.2. Please see the Standard Figure A.2, and the actual output of the MegaPulse DF-1 Lead Connector Tester.
A front-panel switch reverses the output polarity. Additionally provided with BNC oscilloscope output for voltage (100:1) and current (1V/mA) for output monitoring.
The waveform output is 1.5uSec rise time, with a 18mS duration, with an output impedance of 500 ohms.
Click for a view of the outputs showing the Current and Voltage BNC jacks.
MegaPulse DF-1 Lead Connector Tester ships complete, ready to perform the DF-1 test:
MegaPulse DF-1 Lead Connector Tester.
Designed to generate the waveform described in AAMI DF-1, Figure A.2., for the Lead Connector electrical isolation test in AAMI DF-1, Annex A.
TOP
Option: 240-240V Mains power (requires 3U enclosure)
Option: IG-Optional IGBT Switch
Option IFD:Insulation Fault Detection.
Option MPH: Monophasic 200uF waveform addition...
Option P6015: High accuracy Tektronix Probe for proper waveform measurement