Defib-5

Defib Surge Tester

DF-1

DF-80

360J

VIDEO : Defib-5 Setup using the Pass/Fail Reference
The MegaPulse Defib-5 Pass - Fail Reference is also available. This reference is provided with high voltage receptacles and jumper cables to connect between the OUTPUT, RETURN, and MEASURING POINT receptacles of the Defib-5 and simulates a borderline passing result (1 volt appearing on the Y1-Y2 jacks +/- 5%) for ensuring tester function before testing an EUT.

The graph above represents a properly functioning Defib-5, which was discharged into the Y1-Y2 measurement points of Table 50. (For a larger representation, click on the thumbnail above.) A functional check of the Defib-5 would be complete by using the Defib-5 Pass - Fail Reference as noted above; then conducting a second test by checking the output of Y1-Y2 with nothing connected to the Defib-5 Measuring Point, resulting in zero volts output, as shown in the graph below.

This reference is provided with high voltage receptacles and jumper cables to connect between the OUTPUT, RETURN, and MEASURING POINT receptacles of the Defib-5.
TOPOption: None