You can download our product manuals in P.D.F format here.

PT-600
MINI PT-600
The PT-600 is designed to perform the 60 Hz power line cross test/AC power fault tests required in BELCORE 1089, UL 1950 and ITU-T Standards for telecommunications equipment.
Models are available in single or dual channel for 2 and 4 wire testing, and with channel outputs of 40 amps for UL Testing, or 60 amps for BELCORE testing, per channel output. Single and dual channel versions are available.
The PT-600 has easy to use controls and output terminals making test set up simple and efficient. The Preset and Adjustable Time Settings and the Preset and Variable Resistors provide for versatility in performing a wide variety of tests.
Measuring Terminals including Voltage Sense and Current Shunt portals provide a safe location for users to accurately measure test voltages and output current while setting up tests. 5-way output terminals are available for the various source impedance requirements making test set up straightforward and hassle-free. The PT-600 has many safety features: front and rear door interlock switches, emergency stop, and fuse protected output terminals for added safety.
Calibration of the PT-600 is designed to be performed with a simple handheld multi-meter and the removable Currents Shunts can be easily sent to a calibration laboratory, if required.
Computer controlled operation is available with the TestMinder PT option.
If your product requires compliance only with certain tests of the standars, you can take a look for the mini PT-600 model for reference of our flexibility and for more information or a quotation, please call or email Jeff Lind at Compliance West USA.
(800) 748-6224, JLind@compwest.com.
Test conduced by the Base models are: |
|
UL 1950 Annex NAC |
|
| (All models) All M-test | |
| (All 2-Channel PT-600's) All M-,L- and F- Test excep Test L5. | |
Belcore GR-1089 (Model PT-600-480-2-60 only) |
|
| Table 4.7, first level power fault test, test 1,2,3,6,& 7 | |
| Table 4.8 Second-Level AC power fault test,Test 2,3 & 4 (except note 8) | |
Belcore GR-1089 issue 6 |
|
| Table 4.4 First level power fault , test 2,2.1,2.2(Is < 1A), 3, 3.1, 3.2, 3.3 (Is< 1A) & 6 | |
| Table 4.5 Second level power fault, Test 3, 4, 4.3, 6, 7, 8, 8.1 & 8.6. | |
Optional test |
|
UAL 1950 Annex NAC |
|
| Test L5 | |
Belcore GR-1089 (Model PT-600-480-2-60 only |
|
| Clause 4.5.11 | |
| Clause 4.5.14 | |
| Clause 4.5.15, Condition 1 and condition 2 | |
| Clause 4.5.16 | |
| Table 4.7, Test 4, 5, 8 & 9 | |
| Table 4-8, Tests 1, 4 (note 8) & 5. | |
ITU-T Tests (Model PT-600-480-2-60 only): |
|
Power Induction and Earth Potential Rise. For tests conducted, and reference to ITU-T Standards, click here.
For diferent or specific test that is dont listened, please give us a call
Option TMP: TestMINDER TestMinder PT automation software.