The MegaPulse Defib Surge is designed to conduct the test shown in EN 45502-1:1997 Sec. 20, "Protection of the device from damage caused by external defibrillators". In accordance with this Standard reference, the Defib Surge is built to the schematic shown in EN 45502-1 Figure 1 using the circuit values shown in Para. 20.1. The output of the Defib Surge is 140 volts peak. A front-panel switch reverses the output polarity which is required for proper testing. The output of the Defib Surge can be measured with an oscilloscope using an appropriate 10X probe. There is no BNC oscilloscope output on the Defib Surge. The 300 ohm resistor used to connect the device is external to the Defib Surge and is not included. The MegaPulse Defib Pulse is additionally provided with a BNC signal generator input for injection of a 5V 10Hz peak to peak signal in accordance with EN 60601-2-27 Para. 51.101, which is used to judge compliance of the Active Implantable Medical Device function as referenced in EN 45502 Para. 20.1.
The MegaPulse Defib Surge uses IGBT switches to deliver a clean signal output.
The Tests contemplated in EN 45502 require 3 positive going pulses separated by 20 seconds; then after a 60 second pause, three negative going pulses separated by 20 seconds. The positive and negative pulses can be delivered manually using the front panel buttons, or the pulses can be set to operate under computer control using the TestMinder Megapulse option.

MegaPulse Defib Surge ships with cables, graphs of theoretical and actual waveforms, and a Calibration Certificate. One-year warranty. Options include 240V input voltage, a PLC Interface Relay Board and computer control via TestMinder MegaPulse.

MegaPulse Defib Surge Tester.

    TOP

Use

Designed to conduct the test shown in EN 45502-1:1997 Sec. 20, "Protection of the device from damage caused by external defibrillators".

    TOP

Option RI: PLC Interface Board
Option TM: TestMINDER automation software
Option P6015: High accuracy Tektronix probe for proper waveform measurement
Option: 240V Mains power

    TOP
Generates the waveform required by EN 45502-1, Sec. 20 for testing of active implantable medical devices, "Protection of the device from damage caused by external defibrillators".
Designed with operator safety in mind: Set up the EUT to the right and behind the tester, and a right-handed operator will find himself standing to the left of the tester, allowing maximum separation.
Theoretical graphs and actual graphs showing the output of your tester are included with each MegaPulse Defib Surge, allowing easy sanity checks and proof to visiting agency engineers of test setups.
Diode-protected BNC Input for signal generator allows verification of proper operation of the AIMD after the test in accordance with EN 45502-1, Para. 20.1, second subparagraph.
Reverse polarity waveforms available by pushing a button on the front panel. No need to change the test setup. Positive polarity indication minimizes errors.
Front panel voltage meter shows residual voltage in charging circuit for safer operation.
1-year calibration cycle.
All test leads included.
1 year warranty