Compliance West USA Hipot Testers Header

MegaPulse DF80 DSW Surge Tester

DF80  image

The MegaPulse DF80 DSW Surge Tester is designed to conduct the test shown in AAMI DF80, App. DD, Table US.DD.1 and Figure US.DD.1 - Damped sinusoidal waveform parameters. The resulting waveform is in accordandance with those requirements. For waveform and tolerance/compliance graphs and information, click here.

The DF80 DSW is built with 50 ohms output impedance. If another impedance is desired, please contact us.

The DF80 DSW is designed as a laboratory instrument and can handle a duty cycle of 1 hit every 5 minutes continuous within tolerances. Other duty cycles are optionally available:

Option 60. Allows one test every 60 sec continuously.
Option 25. Allows one test every 25 sec continuously.
Option 12. Allows one test every 12 sec continuously.

The power output of the DF80 DSW is 360J. A front panel switch reverses the output polarity. The output of the Defib Surge can be measured with an oscilloscope using an appropriate 10X probe. There is no BNC oscilloscope output on the Defib Surge.

Positive and negative pulses can be delivered manually using the front panel buttons, or the pulses can be set to operate under computer control using the TestMinder Megapulse option.

MegaPulse DF80 DSW ships with cables, graphs of theoretical and actual waveforms, and a Calibration Certificate. One-year warranty. Options include 240V input voltage, a PLC Interface Relay Board, a computer control via Testminder MegaPulse and the duty cycle options described above.

MegaPulse DF80 DSW Tester.

    TOP

Use

Designed to conduct the test shown in AAMI DF80, App. DD, Table US.DD.1 and Figure US.DD.1 - Damped sinusoidal waveform parameters.

    TOP

Option RI: PLC Interface Board
Option TM: TestMINDER automation software
Option P6015: High accuracy Tektronix probe for proper waveform measurement
Option: 240V Mains power

    TOP
Generates the waveform required by AAMI DF 80 App. DD, Specifications for damped sinusoidal output waveforms.
Designed with operator safety in mind: Set up the EUT to the right and behind the tester, and a right-handed operator will find himself standing to the left of the tester, allowing maximum separation.
Theoretical graphs and actual graphs showing the output of your tester are included with each MegaPulse Defib Surge, allowing easy sanity checks and proof to visiting agency engineers of test setups.
360J output; 50 ohms impedance standard.
Reverse polarity waveforms available by pushing a button on the front panel. No need to change the test setup. Positive polarity indication minimizes errors.
Front panel voltage meter shows residual voltage in charging circuit for safer operation.
1-year calibration cycle.
All test leads included.
1 year warranty

    TOP

Ticket info - call 800-555-1212