Defib-5 Pass-Fail

Defib-5

Defib Surge Tester

DF-1

360J

The MegaPulse DF80 DSW Surge Tester is designed to conduct the test shown in AAMI DF80, App. DD, Table US.DD.1 and Figure US.DD.1 - Damped sinusoidal waveform parameters. The resulting waveform is in accordandance with those requirements. For waveform and tolerance/compliance graphs and information, click here.
The DF80 DSW is built with 50 ohms output impedance. If another impedance is desired, please contact us.
The DF80 DSW is designed as a laboratory instrument and can handle a duty cycle of 1 hit every 5 minutes continuous within tolerances. Other duty cycles are optionally available:
Option 60. Allows one test every 60 sec continuously.
Option 25. Allows one test every 25 sec continuously.
Option 12. Allows one test every 12 sec continuously.
The power output of the DF80 DSW is 360J. A front panel switch reverses the output polarity. The output of the Defib Surge can be measured with an oscilloscope using an appropriate 10X probe. There is no BNC oscilloscope output on the Defib Surge.
The MegaPulse DF80 DSW uses IGBT switches to deliver a clean signal output.
Positive and negative pulses can be delivered manually using the front panel buttons, or the pulses can be set to operate under computer control using the TestMinder Megapulse option.
MegaPulse DF80 DSW ships with cables, graphs of theoretical and actual waveforms, and a Calibration Certificate. One-year warranty. Options include 240V input voltage, a PLC Interface Relay Board, a computer control via Testminder MegaPulse and the duty cycle options described above.
MegaPulse DF80 DSW Tester.
Designed to conduct the test shown in AAMI DF80, App. DD, Table US.DD.1 and Figure US.DD.1 - Damped sinusoidal waveform parameters.
TOP
Option RI: PLC Interface Board
Option TM: TestMINDER automation software
Option P6015: High accuracy Tektronix probe for proper waveform measurement
Option: 240V Mains power