Defib-5 Pass-Fail

Defib-5P

Defib Surge Tester

DF-1

360J

The MegaPulse DF-80P DSW Surge Tester is designed to conduct the test shown in AAMI DF80, App. DD, Table US.DD.1 and Figure US.DD.1 - Damped sinusoidal waveform parameters. The resulting waveform is in accordandance with those requirements. For waveform and tolerance/compliance graphs and information, click here.
The DF-80P DSW is built with 50 ohms output impedance. If another impedance is desired, please contact us.
The DF-80P DSW is designed as a laboratory instrument and can handle a duty cycle of 1 hit every 5 minutes continuous within tolerances. Other duty cycles are optionally available:
Option 60. Allows one test every 60 sec continuously.
Option 25. Allows one test every 25 sec continuously.
Option 12. Allows one test every 12 sec continuously.
The power output of the DF-80P DSW is 360J. A front panel switch reverses the output polarity. The output of the Defib Surge can be measured with an oscilloscope using an appropriate 10X probe. There is no BNC oscilloscope output on the Defib Surge.
Positive and negative pulses can be delivered manually using the front panel buttons, or the pulses can be set to operate under computer control using the TestMinder Megapulse option.
MegaPulse DF-80P DSW ships with cables, graphs of theoretical and actual waveforms, and a Calibration Certificate. One-year warranty. Options include 240V input voltage, a PLC Interface Relay Board, a computer control via Testminder MegaPulse and the duty cycle options described above.
| Maximum Voltage output | 5000 V |
|
| Meter accuracy | ±3 % |
|
| Operation temperature | 15-40 °C |
|
| Relativity humidity range | 0-90 % Non condensing |
|
| Standards | AAMI DF80, App. DD, |
|
Table US.DD.1 and Figure |
||
US.DD.1 |
||
| Storage | No |
|
| Duty Cicle | 1 pulse every 5 minutes |
|
| Shipping weight | 30 Lbs |
|
| Product weight | 28 Lbs |
|
| Warranty | 1 Year |
|
| IP rating | N/A |
|
| Calibration | 1 Year |
|
| ECCN number | 3B992 |
|
| HTS/Schedule B | 9030.39.0100 |
|
| Product Dimensions | 16.8 W x 16.8 D x 5.5 H in |
|
| UPC | N/A |
|
| PLC interface | Optional |
|
| Autotriger | Optional |
|
| Continuos Full voltage | Optional |
|
| Testminder | Optional |
|
| Energy calculation | N/A 360 J |
|
| Power supply voltage | 120 VAC 50/60 Hz |
|
| what is in the box | Megapulse DF-80P Tester |
|
2 High voltage tes leads |
||
18 AWG Line power cord |
||
Instruction manual |
||
| Interfaces I/O | USB optional |
|
| Price | Call |
|
TOP
Option RI: PLC Interface Board
Option TM: TestMINDER automation software for P series
Option P6015: High accuracy Tektronix probe for proper waveform measurement
Option: 240V Mains power
Option 60. Allows one test every 60 sec continuously.
Option 25. Allows one test every 25 sec continuously.
Option 12. Allows one test every 12 sec continuously.